Bidhan Chandra Krishi Viswavidyalaya Tender

Search from over 10,000+ business opportunities that have been announced from the biggest West Bengal tenders, e tender West Bengal, WB government tenders, and the Online Tender Notices posted on West Bengal e tenders to date. Find opportunities that have come in from multiple corporations, West Bengal government departments, State PSUs, and several Private companies from the state.

Bidhan Chandra Krishi Viswavidyalaya - BCVK Tender

Goods
Laboratory Equipment and Services
Eprocure
Opening Date14 Jul 2022
Closing Date29 Jul 2022
Tender AmountRefer Documents 
Notes
Download

Costs

Description

Unlock the Tender details for free
Scription for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user this is tender description for unregistered user

Contact

Unlock the Tender details for free
GEM & Bid Advisory Services
Get portal registration, tender bidding, product/service listing or vendor/MSME certification services at a nominal cost

BOQ Items

... More
DO_NOT_SHOW

-

-

-

View all details for FREE
... More
Dryer:-Inner dimensions : 36" x 36" x 24", double door, tray size: 16" X 22" X 1", LOAD : 8 KW No of trays-24 Trays -Construction : The drier should be double walled in construction. The Inner chamber should be made of Mild Steel Sheet and the outer is made of C.R.C steel sheet, with hummerton grey paint. Fabricated rigid iron frame. The joints of the inner chamber should be electrically wielded. -INSULATION : The space between the two walls should be adequately insulated with best quality fiber glass wool to promote efficiency as well as to minimize heat loss during the operation of the Drier. -Door : The Door of the Drier should also be double walled in construction and insulated as well with fiber glass wool. The inner and outer side of the door should be made of M.S Sheet, finished internally with high registance alluminium paint and externally with hummerton grey paint. -Heating element & temperature control : 'A' grade, 'U' shaped tubular heaters, to attain high temperature should be fixed on the two sides of the working chamber. The Temperature shall range from 50 C plus ambient to 250° C ± 5º C. The functioning of the thermostatic control should be indicated by a lamp. -Air circulation : The Drier should be fixed with 1 Nos. of 1 H.P. Motor pulley driven Blowers to produce a rapid movement of a large volume of air over the heating elements, which pass through the air ducts of the inner chamber and should be re-circulated uniformly all over the chamber, maintaining uniform temperature within the working area. -Control unit A separate Control Unit should be provided on the right side of the Drier, which houses an Air - Break Contactor & a DOL starter for the motor Blower, Mains and Load Indicating Lamps, thermostat, a Fine Control Knob, H.R.P Safety Fuse and an 'ON' - 'OFF' Switch for the Heating Elements . Warranty: Three year full replacement warranty

1

-

-

... More
Weighing Balance:Analytical Specifications : - Maximum Capacity : 220 g - Readability : 0.1 mg - Minimum Weight (USP), Typical : 160 mg - Repeatability : 0.08 mg - Weighing Pan Diameter : 90 mm - Adjustment : Internal - Interaces : RS 232 - Dimensions (DxHxW) : 344 mm x 344 mm x 210 mm - Housing : Die-cast aluminum, plastic ABS - Linearity : 0.2 mg - Linearity : 0.1 mg - Legal for Trade : No. - Readability (Certified) : 0.1 mg - Hygenic Design : Yes, Warranty: Three year full replacement

1

-

-

... More
Digital pH meter:pH Instruments include Operating Manual Power Adapter, also with LE 410 pH electrode, electrode arm, 2 x pH 4.01 and 2 x pH 7.00 buffer sachets. SPECIFICATION: ● Measuring Range:-2.00…16.00 pH, -2000…2000mv,-5°C…105°C ● Resolution: 0.01pH/0.1pH, 1mv, 0.1°C ● Error Limits: ±0.01 pH, ±1 mv, 0.3°C ● Calibration: 5 points/ 4 predefined buffer groups. ● Power Supply: 100-240 V / 50-60 Hz, 12V DC ● Size: 227 x 147 x 70 mm/0.63 kg ● Display: 4.3” Segmented Liquid Crystal ● Temperature Compensation: ATC & MTC ● Input: BNC, CINCH, NTC 30k Ω ● Data Storage: 200 Measurements, Current calibration. ● Connectors: RS232 &USB – Interface Warranty: Three years full comprehensive warranty from date of installation

1

-

-

... More
Centrifuge:- Control : Microprocessor with Variable Frequency Drive - Max. Speed : 20000 RPM - Max. RCF : 37000 x g - Choice of Tube Size : 0.2 ml to 100 ml - Max. Capacity : 400 ml - Motor : Brushless Induction Motor - Display : Digital LCD - Program Features : User defined RPM / RCF, Acceleration, Deceleration & Timer with 25 Programs. - Acceleration / Deceleration : 12 - 510 Secs - Timer : 1 - 120 Min - Brake : Dynamic Brake - Safety : Sturdy Cabinet, Lid Lock & Imbalance Protection, Gas Spring - Power Supply: 230 V, 50 Hz. - with 8 x 50ml angle rotor with lid, polypropylene tubes Warranty: Three years full comprehensive warranty from date of installation

1

-

-

... More
Mechanical Rotary Shaker:Type- Double Decker mechanical rotary shaker The Rotary Shaker must be designed for continuous, Heavy Duty, smooth and Noiseless operation. -Housed atop made of mild steel duly powder coated. -It is designed to accommodate flasks 100, 250 & 500 mm capacities. -Provided with Speed Regulator for continuously variable speed 60 to 180 RPM -Driven by 1/8 Hp DC Motor. -Electrically operated on 220 / 230 volt AC Main.). -Flask Capacity : 250 ml x 36 Nos. in lower deck and 500 x 25 Nos. in upper deck Tray size : 650mm x 650mm. With rpm meter

1

-

-

... More
Muffle furnace:a. Inside Chamber Size: at least 6" x 6“x 12", 3.6 L approx b. Construction: The Casing or the outer body should be made of thick Mild Steel Sheet. Hinge type side wise Door and Peep hole should be provided. The casing should be finished in Heat & Rust Proof attractive Silver Ash Hammertone spray painting/Heat Resistant Aluminium Paint. c. Heating: Heating Chamber (Muffle) consists of grooved refractories having grooved outer surface on which Kanthal A-1, element in Coil Form should be placed in between the Mild Steel Sheet casing and the Heating Chamber. Hot and Cold faced insulating bricks (Kyanite Base) should be suitably cemented forming a Refractory Lining to keep the shell Temperature within 80°C. (Ceramic Blanket & Ceramic Board can be provided at extra cost). d. Temperature: Maximum temperature range shall be upto 1100°C. This temperature should be reached within 1½ hours provided constant voltage of 220 V supply is ensured. Continuous working Temperature 1000°C ± 10° C temperature uniformity in the chamber should be maintained. e. Controls: Microprocessor based Digital PID Indicator cum Temperature Controller Thermocouple and suitable compensating Cable controls and indicates the temperature. Panel Box: An electrical control panel box consisting of Digital Temperature indicator cum controller, indicating neon lamps for main and load indication, relay, energy regulator, ON/OFF toggle switch & main cord for direct connection to laboratory mains should be provided. f. Power: 220 Volts AC single phase, 50 Hz. Load : 4 KW Warranty: Three years full comprehensive warranty from date of installation

1

-

-



Evaluation Notes How It Works ?

Potential Partner

Select Your Requirements
Disclaimer: BidAssist has made every reasonable effort to ensure that the information is accurate and authentic however it cannot be held liable for any third-party claims or losses of any damages. BidAssist makes no warranty, expressed or implied, as to the results obtained from the use of the information. If you think you may have noticed any error or omission, please let us know by contacting contact@bidassist.com.